![]() Similarly, the system allows the clinician to decide when to turn ON inhalation anesthetic agent during the case. The system allows this flexibility so the clinician can decide which mode of ventilation is appropriate for that phase of the case. ![]() mechanical ventilation by toggling the bag/vent switch from “bag” to “vent.”.manual ventilation by moving the bag/vent switch to “bag” or.When using the GE Healthcare Aisys or Aisys CS2 anesthesia systems and “start case” is selected, the system allows the clinician to begin the case with either: ![]() Thank you for writing about this important issue, which is critical to patient safety during the perioperative period. We welcome a reply from representatives from GE. if the ventilator is in operation, and the VA was turned off due to cassette change, but not turned back on.if the ventilator is in operation, but no volatile anesthetic (VA) or nitrous oxide was turned on.After each of these incidents we reported to GE (we use GE Healthcare Aisys™ anesthesia machines) that we felt there should be an alarm on the machine to alert providers if the following circumstances occur: My hospital had three incidents in the last four years that may be related to volatile anesthetics unintentionally not being delivered to the patient.